How to Clean and Store pH & ORP Electrodes
ASI...MEASURING YOUR SUCCESS!
pH Electrode Storage and Maintenance

Routine Storage

Long-term Storage

Maintenance

Routine Cleaning

Weekly Maintenance

Removal of Membrane/Junction Deposits



pH Electrode Storage and Maintenance

To ensure a quick response and free-flowing liquid junction, the sensing element and reference junction must not be allowed to dry out.  The following instructions apply to refillable electrodes.  For gel-filled electrodes, consult instruction manual.

^ Top

Routine Storage

Soak electrode in a pH Electrode Storage Solution.  If a storage solution is unavailable, pH 4 buffer or 0.1 M KCl may be used.  The fill hole should be covered to prohibit evaporation of reference fill solution.

^ Top

Long-term Storage

Cover sensing element with its protective cap containing a several drops of pH/ORP storage solution.
NOTE: Before returning to use, rinse well with deionized water, refill with appropriate fill solution and soak in pH 4 buffer for 1-2 hours.

^ Top

Maintenance

Cleaning your electrode between and after use will help extend the life of your electrode and avoid the cost of early replacement.

^ Top

Routine Cleaning

General - Soak electrode in 0.1 M HCl or HNO3 for half an hour, followed by soaking it in storage solution for at least two hours.

Reference junction - To clean a fouled junction:

  • Heat electrode in 0.1 M KCl at 70°C (60°C for calomel electrodes) for 10 minutes.
  • Drain and refill with fresh filling solution.
  • Verify junction flow by suspending electrode in air and letting filling solution flow out and crystallize.
  • Soak in storage solution for one hour (24 hours for calomel electrodes) prior to use.
  • ^ Top

    Weekly Maintenance

    Inspect electrodes for scratches, cracks, salt crystal buildup, or membrane/junction deposits.

    Rinse off any salt buildup with distilled water, and remove any membrane/junction deposits as directed in cleaning procedures below.

    The reference chamber should be drained, flushed with fresh filling solution, and refilled.

    ^ Top

    Removal of Membrane/Junction Deposits

  • Grease and oil - Rinse electrode with a mild soap
  • Inorganic - Soak electrodes in 0.1M ??? solution for 10 minutes
  • Protein - Soak in a 0.1M HCl solution for 10 minutes, rinse with DI water
  • After any of these cleaning procedures, flush and refill the reference chamber and soak the electrodes in storage solution overnight.
  • ^ Top